Y-90 Dosimetry Calculator
Single-compartment MIRD and partition-model dosimetry for glass and resin microspheres, with decay correction to the treatment day, microsphere activity, and sphere density in tumor and perfused liver.
Advanced assumptions
Absorbed dose
–
| Compartment | Mass | Dose |
|---|
Sphere density
–
| Compartment | Spheres | Per mL |
|---|
Vial & decay
–
Sphere count vs. treatment day
Same administered activity, delivered on different days relative to calibration.
Method
Decay. Y-90 half-life 64.1 h (λ = ln2 / 2.671 d). Glass activity per sphere falls as the vial ages past calibration; resin treated before calibration carries more activity per sphere.
MIRD single compartment. Activity is assumed uniform across the perfused territory.
Partition model. Liver activity splits between tumor and normal liver according to T/N and compartment mass.
Sphere density. N = A / asphere(t); spheres reaching liver are partitioned with the same activity fractions as the dose and divided by compartment volume. In MIRD mode the tumor/normal split is shown as an estimate using the entered T/N.
- Default per-sphere activity at calibration: glass 4,300 Bq, resin 50 Bq (editable under Advanced).
- Labeling windows flagged: glass up to 12 days post-calibration; resin within 24 h of calibration.
Educational demonstration only. Not validated for clinical treatment planning; verify all prescriptions against the manufacturer's instructions for use and institutional protocols.